Product Overview

P5530 Automatic Trace Moisture Analyzer is suitable for GB/T7600-1987 "Determination of Moisture Content in Running Transformer Oil (Coulomb Method)", SH/T0255-1992 "Determination of Moisture Content in Additives and Lubricating Oil with Additives (Electricity Method)", SH/T0246-1992 "Determination of Moisture Content in Light Petroleum Products (Electricity Method)", accurately measuring trace moisture in liquids, solids, gases, and used in industries such as power, petroleum, chemical, pharmaceutical, and food. It can quickly determine the moisture content of chemical, petroleum, pharmaceutical, pesticide and other products such as alcohols, oils, lipids, ethers, esters, acids, alkanes, benzenes, amines, organic solvents, pesticides, phenols, and pharmaceutical raw materials.

Principle

Iodine is formed on the anode through electrolysis, and the generated iodine is directly proportional to the amount of charge according to Faraday's law. The number of moles of iodine involved in the reaction is equal to the number of moles of water. Inject the sample into the electrolyte, and the water in the sample participates in the reaction. The instrument can reflect the consumption of iodine during the reaction process, and the consumption of iodine can be calculated based on the amount of electricity required to electrolyze the same amount of iodine. After calculation by the instrument, the measured water content can be directly displayed on the liquid crystal display.

Features

*Adopting high-resolution color 480 × 272 touch dot matrix LCD, with a fully Chinese human-machine dialogue interface

*The use of a switch constant current source electrolysis circuit has reduced the power consumption of the instrument

*The high-precision measurement electrode signal generation and detection circuit enables fast and accurate determination of the electrolysis endpoint, and has strong anti-interference ability

*Using two methods of reagent blank current compensation and equilibrium point drift compensation to correct measurement results

*Built in 9 calculation formulas to meet the calculation of moisture content under different measurement conditions

*Can be converted separately or simultaneously to mg/L, ppm, mg/kg,%, reducing measurement frequency and reagent consumption

*Sample volume, mass, and density can be adjusted during or after the testing process

*9-speed mixing speed adjustment; 9-level electrolytic gain adjustment

*Equipped with automatic detection and recording functions for measuring electrode open circuit faults and short circuit faults

*Automatically store historical records with time stamps

*The operation interface can switch between Chinese and English

technical parameters

Measurement range: 3 μ g~100mg

Instrument accuracy: 3 μ g to 1mg ± 3 μ g; More than 1mg is 0.3% (excluding injection error)

Resolution: 0.1 μ g

Electrolytic current: 0-400mA

Electrolysis rate: 2.4mg/min

Communication ports: RS485, RS232

Communication protocol: ModBus RTU

Power consumption of the whole machine: 30W

Power supply voltage: AC220V ± 10%, 50Hz ± 10%

Environmental temperature: (20-30) ℃

Environmental humidity:<70% RH (no condensation)

Dimensions: 320mm × 240mm × 200mm

Instrument weight: 4.7kg