Impact test chamber

★★★★★ 4.9/5 (156 reviews)

Product Overview

The high and low temperature impact test chamber is used to test the degree to which composite materials and material structures can withstand continuous environments of high and extremely low temperatures in an instant, in order to test the chemical changes or physical damage caused by thermal expansion and contraction in the shortest possible time.

II Technical parameters of high and low temperature impact test chamber manufacturer

Inner box size: 50 × 40 × 40 (can be customized according to customer requirements)

Outer box size: 150 × 190 × 160

Test method: Pneumatic air door switch between 2 or 3 greenhouse modes

Preheating temperature range for high temperature chamber:+60~+200 ℃

Heating rate:+60 →+200 ℃ for about 20 minutes

Low temperature chamber pre cooling temperature range: -78 → 0 ℃

Cooling rate:+20 → -75 ℃ for about 70 minutes

Laboratory temperature deviation: ± 2 ℃

Temperature range: A: -40 ℃~+150 ℃, B: -55 ℃~150 ℃, C: -65 ℃~+150 ℃

Temperature recovery time: within 5 minutes

Material: High density cold-rolled steel plate with nano spray coating for the shell; Stainless steel (SUS304) inside, polyurethane foam thermal insulation material

Heater: Nickel chromium alloy electric heater

Fan: shared centrifugal fan for high temperature and environmental temperature exposure, axial flow fan for preheating

Heater and cooler: Nickel chromium alloy electric heating, finned cooler, and cold storage device

Fan: Centrifugal fan

Pneumatic cylinder: used for driving various dampers during high temperature, ambient temperature, and low temperature exposure

Air Compressor: Provides compressed air to drive pneumatic dampers (optional)

Refrigeration method: mechanical compression binary cascade refrigeration

Refrigeration compressor: semi enclosed piston type

Refrigerant: 23/R404 environmentally friendly

Condenser: Stainless steel brazed plate heat exchanger

Operation interface: 6 "color LCD display touch screen, Chinese menu prompts

Program memory capacity: 100 sets of user programs (customizable and modifiable)

Set indicator range: time: 1 minute~99 hours and 59 minutes, cycle: 1~999 cycles

Resolution: Temperature: 0.01 ℃, Time: 1 minute

Power supply: AC380 ± 10% V 50 ± 0.5Hz, three-phase four wire+protective ground wire

Power (kw): 20, 21, 22

Cooling circulating water: water pressure: 0.2~0.4MPa, water temperature: ≤ 30 ℃

Compressed air: 0.5~0.7MPa

III High and low temperature impact test chamber manufacturer's safety device

Leakage circuit breaker, high and low temperature protection device (built-in controller), high and low temperature protection device, exhaust valve, sample power control terminal (contact capacity: AC 250V 6 amperes), high and low temperature room over temperature protection switch, high and low temperature room over temperature protection (built-in controller), compressor over pressure and over temperature protection, water cut-off relay, fan thermal relay, motor temperature switch, motor reverse prevention relay, compressed air pressure switch, fuse, external alarm terminal (contact capacity: AC 250V 2 amperes)